10 other products:
Ningbo David TI-2000 Transport Incubator
Closed transport incubator with servo controlled air and skin temperature, includes 2 oxygen cylinders and battery.
Centrífuga INMUNOHEMATOLOGIA BIORIDGE
Microcontroller control, brushless motor, Stable operation, high speed precision, Touch panel, program recording, LED display, Easy interface, RPM or RCF speed conversion.
Electric Bed 5 Movements Saikang SKD-A
5-movement electric bed with 4 side rails, Linak motors, 220 kg capacity.
Laryngoscope Shells
Pre Hospital Care, Diagnostic Equipment. Welch Allyn kits, leaflet, curved leaflet, curved leaflet with fiber, laryngoscope leaflet, straight leaflet, straight leaflet with fiber, Welch Allyn leaflets.
Camilla Divan Estructura Tubular En Aluminio
Stretcher divan, in tubular aluminum structure, upholstered in 3 cm foam, lined in cordoban, easy to wash and asepsis.
Ambu Cervical Neck Immobilizer Adult Adjustable Perfit Ace
Adjustable 16 Size Cervical Immobilizer (Neck Brace) for Adult Patient
Randy Nasco 5'5 "105 Pound Rescue Mannequin
Rescue Randy was developed for realistic training of handling, transporting and extracting adult or juvenile victims. These mannequins can be safely used in situations that are too dangerous or uncomfortable for human volunteers.
PARAMED'S Tactical Sanitary Life Support Backpack
Backpack specially designed for emergency and rescue workers that will allow to organize in a comfortable way all the essential material and instruments for rescues and emergency interventions in mountains and areas of difficult access. This...
EXTREME'S Emergency Bag Basic Life Support
Its internal honeycomb organization allows better organization of content. It has removable compartments of different colors.
Product Comments





Descripción
Brand: Mindray
Model: BeneHeart D3
Origin: China
Features:
• 4 in 1 design.
• Manual defibrillation, AED, pacemaker and monitoring.
• 7 ”TFT color screen.
• Defibrillation, synchronized cardioversion and AED.
• Administration of doses from 1J to 360J.
• Compact design - easy to transport and use.
Technical specifications:
Defibrillator
• Waveform: Biphasic Truncated Exponential (BTE) waveform, with impedance compensation.
• Accuracy: ± 2 J with 15% adjustment, For any value below 50 Ohm.
• Charging time: Less than 5 seconds for 200 joule, with maximum load. Less than 8 seconds for 360 joule, full load.
• Patient impedance: 20 to 200 Ohm.
AED mode
• Output: configured by user.
• AED shock series: level: 100 to 360 J, configurable Shock series: 1, 2, 3 configurable.
Screen
• Type: LCD, color TFT.
• Dimensions: 7 ”.
• Resolution: 800x600 pixels.
• Waves on screen: 3 Channels.
• Sweep Time: 16s Max. (ECG).
Non-invasive pacemaker
• Waveform: Square pulse wave.
• Pulse width: 20 ms, ± 5%.
• Reset period: 200 to 300 ms, ± 3%.
• Rhythm Mode: By demand or arrangement.
• Pace range: 40 bpm to 170 bpm, ± 1.5%.
ECG monitoring
• Cable Type: 3 Lead ECG, 5 Lead ECG, PADS / PADDLES.
• Selection: I, II, III, aVL, aVF, V, Pads / Paddles.
• Sample range: Adult: 15 to 300 bpm Pediatric: 15 to 350 bpm Neonatal: 15 to 350 bpm.
• Resolution: 1 bpm.
• Arrhythmia: yes.
• Alarm: yes.
Printer
• Method: Thermal printer (high resolution).
• Waveform: Max. 3 Channels.
• Speed: 25 mm / s, 50 mm / s.
• Paper: 50 mm.
• Reports: You can print: Event summary, data tabulation, Waveform.
Data storage
• Patient file: Max. 100 patients.
• Events: 1000 events per patient.
• Waveform: 24 hours continuous ECG.
Supplied Accessories:
• Built-in printer.
• 1 rechargeable Li-ion battery.
• 3 Rolls of thermal paper.
• 1 Kit of external pallets.
• 1 5-lead ECG cable.
• 1 Electrode gel.
• 1 Power cord.
• User manuals.
Información Técnica
Dimensions | 28.8cm (w) x 20.3cm (d) x 27.5cm (h) |
Weight | 4.7 kg with all modules 0.5 kg batteries 0.8 external paddles |
Physical and environmental requirements: | |
Waterproof | IPx4 (No external power connection). IPx1 (With external power connection) |
Solid resistance | IPx4 |
Operating temperature | 0 ºC - 45 ºC |
Storage temperature | -30 ºC - 70 ºC |
RH | 10% - 95% non-condensing |
Altitude | 381 m to +4575 m |
Shock and vibration | Meets requirement 21.102, ISO9919 (Shock and vibration). Meets requirement 6.3.4.2, EN1789 (Use in ambulances). |
Freefall | Meets requirement 6.4.2, EN1789 (Drops of: 0.7 m). |
EMC | IEC60601-1-2 |
Security | EN / IEC60601-1 |
Screen: | |
Guy | LCD, color TFT |
Dimensions | 7 " |
Resolution | 800 x 480 pixels |
Waves on screen | 3 channels |
Sweep time | 16 s max. (ECG) |
Power supply: | |
Entrance | AC |
Line voltage: | AC 100 to 240 V (± 10%) |
Stream | 0.8 to 1.8 A |
Frequency | 50 / 60Hz (± 3Hz) |
DC input | AC-DC converter |
Input voltage | DC 12V |
Consumption | 90 W |
Battery: | |
Guy | 3 Ah, 14.8V, rechargeable Lithium ion |
Quantity | Max. 1 |
Loading time | 2 hours at 80% load. 3 hours at 100% with the device off. |
Functioning | More than 100 shocks at 360 J |
Printer: | |
Method | Thermal printer (high resolution). |
Waveform | Max. 3 channels |
Speed | 25mm / s, 50mm / s |
Paper | 50 mm |
Reports | Summary. of events Data tabulation Waveform. |
Data storage | |
Patient file | 100 patients |
Events (edit) | 1000 events per patient |
Waveform | 24 hours continuous ECG |
Data export | They can be exported to PC or USB Flash memory |
Defibrillator: | |
Waveform | Truncated Biphasic waveform. |
Accuracy | ± 2 J with 15% adjustment, for any value below 50 Ohm. |
Loading time | Less than 5 s for 200 J, with maximum load. Less than 8 s for 360 J, with maximum load. |
Patient impedance | 20 to 200 Ohm |
Manual mode: | |
Departure | : 1,2,3,4,5,6,7,8,9,10,15,20,30,50,70,100,150,170,200,300,360 J |
Synchronized cardioversion | The transfer begins 60 ms from the QRS peak. The transfer begins with 25 ms of the external sync pulse. |
AED mode: | |
Departure | User configured |
DEA download series | Level: 100 to 360 J, configurable. Download series: 1, 2, 3 configurable. |
Non-invasive pacemaker: | |
Waveform | Square wave pulse |
Pulse width | 20 ms, ± 5% |
Reassembly period | 200 to 300 ms, ± 3% |
Rhythm mode | On demand or fixed |
Rhythm range | 40 ppm to 170 ppm, ± 1.5% |
ECG monitoring: | |
Type of cable | 3 Lead ECG, 5 Lead ECG, PADS / PADDLES |
Selection | I, II, III, aVL, aVF, V, Pads / Paddles |
Range sample | Adult: 15 to 300 bpm. Pediatric: 15 to 350 bpm. Neonatal: 15 to 350 bpm |
Resolution | 1 bpm |
Arrhythmia | Yes |
Alarm | Yes |
Breathing: | |
Technique | Trans-Thoracic Impedance |
Rank | Adult: 0 to 120 rpm: Pediatric, Neonatal: 0 to 150 rpm |
Resolution | 1 rpm |
Standard equipment: | |
With pacemaker | 1 AC power cord. 1 Internal printer with 1 paper. 1 Rechargeable Lithium Battery. 1 Pallet Kit. 1 Kit of Electrodes and ECG Leads. 1 Def / Pacemaker / AED Lead. 1 Def / Pacemaker / AED Pads |
No pacemaker | 1 AC power cord. 1 Internal printer with 1 paper. 1 Rechargeable Lithium Battery. 1 Pallet Kit. 1 Kit of Electrodes and ECG Leads |