10 other products:
Magill Dimeda forceps
Through the mouth, size 17 cm for neonates, 20 cm for children and 25 cm for adults.
Ambu Cervical Neck Immobilizer Adult Adjustable Perfit Ace
Adjustable 16 Size Cervical Immobilizer (Neck Brace) for Adult Patient
Diagnostic Equipment Sense Organs Welch Allyn Pocket Plus LED
Case of sense organs (otoscope and ophthalmoscope) in soft case, available in onyx, lilac, snow and indigo colors.
Ambu® Silicone Oval Resuscitator Reusable Resuscitator
The Ambu® Oval Silicone Resuscitator has been designed for manual ventilation, from neonatal to adult patients. It is a reusable silicone resuscitator. The texture of the surface of the silicone resuscitators, combined with the grip handle...
Respirator N95 M920 Particulate Material Steelpro
Thermo-formed and heat-sealed maintenance-free particulate respirator, designed to achieve greater comfort while providing adequate protection for the user. It allows the filtration of oil-free particulate matter down to a size of 0.3 μm...
Lectern Chrome Plated 2 Wheels Mafet M-005
Lectern with or without wheels, stainless steel, chromed, painted or ABS base, model of choice.
Antiseptic Alcohol 70% INVIMA Gallon / 120ml / 350ml / 700ml / 3750ml
It is used for the disinfection of the skin, utensils, equipment and laboratory material. For direct use. (It also serves in skin antisepsis). Useful for disinfection in clinics, hospitals, laboratories, industries and companies in general....
Camilla Divan Estructura Tubular En Aluminio
Stretcher divan, in tubular aluminum structure, upholstered in 3 cm foam, lined in cordoban, easy to wash and asepsis.
Product Comments
image Featured Descripción
Brand: Mindray
Model: BeneHeart D3
Origin: China
Features:
• 4 in 1 design.
• Manual defibrillation, AED, pacemaker and monitoring.
• 7 ”TFT color screen.
• Defibrillation, synchronized cardioversion and AED.
• Administration of doses from 1J to 360J.
• Compact design - easy to transport and use.
Technical specifications:
Defibrillator
• Waveform: Biphasic Truncated Exponential (BTE) waveform, with impedance compensation.
• Accuracy: ± 2 J with 15% adjustment, For any value below 50 Ohm.
• Charging time: Less than 5 seconds for 200 joule, with maximum load. Less than 8 seconds for 360 joule, full load.
• Patient impedance: 20 to 200 Ohm.
AED mode
• Output: configured by user.
• AED shock series: level: 100 to 360 J, configurable Shock series: 1, 2, 3 configurable.
Screen
• Type: LCD, color TFT.
• Dimensions: 7 ”.
• Resolution: 800x600 pixels.
• Waves on screen: 3 Channels.
• Sweep Time: 16s Max. (ECG).
Non-invasive pacemaker
• Waveform: Square pulse wave.
• Pulse width: 20 ms, ± 5%.
• Reset period: 200 to 300 ms, ± 3%.
• Rhythm Mode: By demand or arrangement.
• Pace range: 40 bpm to 170 bpm, ± 1.5%.
ECG monitoring
• Cable Type: 3 Lead ECG, 5 Lead ECG, PADS / PADDLES.
• Selection: I, II, III, aVL, aVF, V, Pads / Paddles.
• Sample range: Adult: 15 to 300 bpm Pediatric: 15 to 350 bpm Neonatal: 15 to 350 bpm.
• Resolution: 1 bpm.
• Arrhythmia: yes.
• Alarm: yes.
Printer
• Method: Thermal printer (high resolution).
• Waveform: Max. 3 Channels.
• Speed: 25 mm / s, 50 mm / s.
• Paper: 50 mm.
• Reports: You can print: Event summary, data tabulation, Waveform.
Data storage
• Patient file: Max. 100 patients.
• Events: 1000 events per patient.
• Waveform: 24 hours continuous ECG.
Supplied Accessories:
• Built-in printer.
• 1 rechargeable Li-ion battery.
• 3 Rolls of thermal paper.
• 1 Kit of external pallets.
• 1 5-lead ECG cable.
• 1 Electrode gel.
• 1 Power cord.
• User manuals.
Información Técnica
| Dimensions | 28.8cm (w) x 20.3cm (d) x 27.5cm (h) |
| Weight | 4.7 kg with all modules 0.5 kg batteries 0.8 external paddles |
| Physical and environmental requirements: | |
| Waterproof | IPx4 (No external power connection). IPx1 (With external power connection) |
| Solid resistance | IPx4 |
| Operating temperature | 0 ºC - 45 ºC |
| Storage temperature | -30 ºC - 70 ºC |
| RH | 10% - 95% non-condensing |
| Altitude | 381 m to +4575 m |
| Shock and vibration | Meets requirement 21.102, ISO9919 (Shock and vibration). Meets requirement 6.3.4.2, EN1789 (Use in ambulances). |
| Freefall | Meets requirement 6.4.2, EN1789 (Drops of: 0.7 m). |
| EMC | IEC60601-1-2 |
| Security | EN / IEC60601-1 |
| Screen: | |
| Guy | LCD, color TFT |
| Dimensions | 7 " |
| Resolution | 800 x 480 pixels |
| Waves on screen | 3 channels |
| Sweep time | 16 s max. (ECG) |
| Power supply: | |
| Entrance | AC |
| Line voltage: | AC 100 to 240 V (± 10%) |
| Stream | 0.8 to 1.8 A |
| Frequency | 50 / 60Hz (± 3Hz) |
| DC input | AC-DC converter |
| Input voltage | DC 12V |
| Consumption | 90 W |
| Battery: | |
| Guy | 3 Ah, 14.8V, rechargeable Lithium ion |
| Quantity | Max. 1 |
| Loading time | 2 hours at 80% load. 3 hours at 100% with the device off. |
| Functioning | More than 100 shocks at 360 J |
| Printer: | |
| Method | Thermal printer (high resolution). |
| Waveform | Max. 3 channels |
| Speed | 25mm / s, 50mm / s |
| Paper | 50 mm |
| Reports | Summary. of events Data tabulation Waveform. |
| Data storage | |
| Patient file | 100 patients |
| Events (edit) | 1000 events per patient |
| Waveform | 24 hours continuous ECG |
| Data export | They can be exported to PC or USB Flash memory |
| Defibrillator: | |
| Waveform | Truncated Biphasic waveform. |
| Accuracy | ± 2 J with 15% adjustment, for any value below 50 Ohm. |
| Loading time | Less than 5 s for 200 J, with maximum load. Less than 8 s for 360 J, with maximum load. |
| Patient impedance | 20 to 200 Ohm |
| Manual mode: | |
| Departure | : 1,2,3,4,5,6,7,8,9,10,15,20,30,50,70,100,150,170,200,300,360 J |
| Synchronized cardioversion | The transfer begins 60 ms from the QRS peak. The transfer begins with 25 ms of the external sync pulse. |
| AED mode: | |
| Departure | User configured |
| DEA download series | Level: 100 to 360 J, configurable. Download series: 1, 2, 3 configurable. |
| Non-invasive pacemaker: | |
| Waveform | Square wave pulse |
| Pulse width | 20 ms, ± 5% |
| Reassembly period | 200 to 300 ms, ± 3% |
| Rhythm mode | On demand or fixed |
| Rhythm range | 40 ppm to 170 ppm, ± 1.5% |
| ECG monitoring: | |
| Type of cable | 3 Lead ECG, 5 Lead ECG, PADS / PADDLES |
| Selection | I, II, III, aVL, aVF, V, Pads / Paddles |
| Range sample | Adult: 15 to 300 bpm. Pediatric: 15 to 350 bpm. Neonatal: 15 to 350 bpm |
| Resolution | 1 bpm |
| Arrhythmia | Yes |
| Alarm | Yes |
| Breathing: | |
| Technique | Trans-Thoracic Impedance |
| Rank | Adult: 0 to 120 rpm: Pediatric, Neonatal: 0 to 150 rpm |
| Resolution | 1 rpm |
| Standard equipment: | |
| With pacemaker | 1 AC power cord. 1 Internal printer with 1 paper. 1 Rechargeable Lithium Battery. 1 Pallet Kit. 1 Kit of Electrodes and ECG Leads. 1 Def / Pacemaker / AED Lead. 1 Def / Pacemaker / AED Pads |
| No pacemaker | 1 AC power cord. 1 Internal printer with 1 paper. 1 Rechargeable Lithium Battery. 1 Pallet Kit. 1 Kit of Electrodes and ECG Leads |
English